Abstract
We design, fabricate and experimentally demonstrate a novel generic method to detect flow rates and precise changes of flow velocity in microfluidic devices. Using our method we can measure flow rates of ~2 mm/s with a resolution of 0.08 mm/s. The operation principle is based on the Doppler shifting of light diffracted from a self-generated periodic array of bubbles within the channel and using self-heterodyne detection to analyze the diffracted light. As such, the device is appealing for variety of “lab on chip” bio-applications where a simple and accurate speed measurement is needed, e.g., for flow-cytometry and cell sorting.
| Original language | English |
|---|---|
| Pages (from-to) | 16799-16807 |
| Number of pages | 9 |
| Journal | Sensors |
| Volume | 14 |
| Issue number | 9 |
| DOIs | |
| State | Published - 10 Sep 2014 |
| Externally published | Yes |
Bibliographical note
Publisher Copyright:© 2014 by the authors; licensee MDPI, Basel, Switzerland.
Keywords
- Micro fluidics devices
- Sensors
- Velocimetry
Fingerprint
Dive into the research topics of 'Doppler-based flow rate sensing in microfluidic channels'. Together they form a unique fingerprint.Cite this