An implantable microfluidic device for self-monitoring of intraocular pressure.,

Araci E. Ismail, Baolong Su, Quake R. Stephen, Yossi Mandel

Research output: Contribution to journalArticlepeer-review

Abstract

Glaucoma is the second most common cause of blindness in the world. It is a multifactorial disease with several risk factors, of which intraocular pressure (IOP) is a primary contributing factor. IOP measurements are used for glaucoma diagnosis and patient monitoring. IOP has wide diurnal fluctuation and is dependent on body posture, so the occasional measurements done by the eye care expert in the clinic can be misleading. Here we show that microfluidic principles can be used to develop an implantable sensor that has a limit of detection of 1 mm Hg, high sensitivity and excellent reproducibility. This device has a simple optical interface that enables IOP to be read with a smartphone camera. This sensor, with its ease of fabrication and simple design, as well as its allowance for IOP home monitoring, offers a promising approach for better care of patients with glaucoma.,
Original languageAmerican English
Pages (from-to)1074-8
JournalNature Medicine,
Volume20
Issue number9,
StatePublished - 2014

Fingerprint

Dive into the research topics of 'An implantable microfluidic device for self-monitoring of intraocular pressure.,'. Together they form a unique fingerprint.

Cite this