TY - JOUR
T1 - A performance study of a new personal neutron dosimetry system at SNRC
AU - Amit, Gal
AU - Even-Hen, Ofir
AU - Awad, Oshrit
AU - Levi, Yaniv
AU - Buchbinder, Lotem
AU - Datz, Hann
N1 - Publisher Copyright:
© The Author(s) 2020. Published by Oxford University Press. All rights reserved.
PY - 2020/7/13
Y1 - 2020/7/13
N2 - This research reviews the performance of a commercial personal fast neutron dosimetry system that has recently been installed in the External Dosimetry Laboratory at Soreq Nuclear Research Center (SNRC). Different kinds of neutron sources like 241Am–Be, 252Cf and some neutron generators are used in the industry as well as in some research institutes and reactors. The workers in these places need to be regularly monitored for neutron radiation. These workers’ individual neutron doses have been estimated at SNRC until recently by a 30-year-old in-house-made system based on the CR-39 ® solid state nuclear track (SSNT) detector technology. Last year, a new Landauer CR-39 ® neutron dosimetry system was adopted, the Neutrak system. This system employs a standard etching technique using NaOH solution, and measures the SSNT dosimeters automatically and in a repeatable manner. This paper reviews the performance of Landauer’s commercial personal neutron dosimetry system. The system performance shows excellent results in terms of linearity, repeatability, robustness, etch parameters sensitivity and detection threshold.
AB - This research reviews the performance of a commercial personal fast neutron dosimetry system that has recently been installed in the External Dosimetry Laboratory at Soreq Nuclear Research Center (SNRC). Different kinds of neutron sources like 241Am–Be, 252Cf and some neutron generators are used in the industry as well as in some research institutes and reactors. The workers in these places need to be regularly monitored for neutron radiation. These workers’ individual neutron doses have been estimated at SNRC until recently by a 30-year-old in-house-made system based on the CR-39 ® solid state nuclear track (SSNT) detector technology. Last year, a new Landauer CR-39 ® neutron dosimetry system was adopted, the Neutrak system. This system employs a standard etching technique using NaOH solution, and measures the SSNT dosimeters automatically and in a repeatable manner. This paper reviews the performance of Landauer’s commercial personal neutron dosimetry system. The system performance shows excellent results in terms of linearity, repeatability, robustness, etch parameters sensitivity and detection threshold.
UR - http://www.scopus.com/inward/record.url?scp=85088487412&partnerID=8YFLogxK
U2 - 10.1093/RPD/NCAA036
DO - 10.1093/RPD/NCAA036
M3 - ???researchoutput.researchoutputtypes.contributiontojournal.article???
C2 - 32318705
AN - SCOPUS:85088487412
SN - 0144-8420
VL - 189
SP - 242
EP - 252
JO - Radiation Protection Dosimetry
JF - Radiation Protection Dosimetry
IS - 2
ER -