A fluorometric-reflectometric technique for simultaneous measurement of ph and motility in ram semen

D. Bar-Sagie, A. Mayevsky, B. Bartoov

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

An optical technique for simultaneous assessment of pH and motility in ram semen was described. Motility was evaluated by the measurement of changes in intensity of reflected light from semen samples illuminated with 366 nm light. pH was measured by using a pH-sensitive fluorescent indicator (umbelliferone). From the changes in fluorescence intensity (450 nm) as a function of the pH of the semen, a calibration curve was constructed. The relationship between pH changes and motility in various metabolic conditions were described. Different kinetics for pH changes occurred when sperm motility was driven by both mitochondrial respiration and fructolysis or by either one of them. An analysis of the fluorescence signal from extracellular or cellular fractions of semen suggested that the pH measurement reflected extracellular pH.

Original languageEnglish
Pages (from-to)27-33
Number of pages7
JournalSystems Biology in Reproductive Medicine
Volume7
Issue number1
DOIs
StatePublished - Aug 1981

Keywords

  • Fluorescence
  • Motility
  • PH
  • Spermatozoa
  • Umbelliferone

Fingerprint

Dive into the research topics of 'A fluorometric-reflectometric technique for simultaneous measurement of ph and motility in ram semen'. Together they form a unique fingerprint.

Cite this