Equipments Details
Description
Tecan infinite M1000 Pro
Booking
Description
Absorption, fluorescence, luminescence, and alpha- screen. The Infinite M1000 PRO offers broad wavelength selection, from UV to NIR, with flexible bandwidth adjustment for optimization of fluorescence intensity based applications.
Features/Specifications
Wavelength selection
Premium Quad4 Monochromators
Absorbance Fluorescence
Bandwidth Fixed, 5 nm Adjustable = 300 nm: 2.5 – 10 nm;
> 300 nm: 5 – 20 nm
Wavelength accuracy ± 0.3 nm = 300 nm: ± 0.5 nm;
> 300 nm: ± 1 nm
Wavelength reproducibility ± 0.3 nm = 300 nm: ± 0.5 nm;
> 300 nm: ± 1 nm
Wavelength range
Absorbance Fluorescence FP
230 – 1,000 nm Ex: 230 – 850 nm; Em: 280 – 850 nm Ex (LED):
470, 530, 590, 635 nm
Em (Mono):
280 – 850 nm
Detectors
Absorbance UV Silicon photodiode, 4-channel parallel reading
Fluorescence (incl. FP) Extended wavelength (UV and far red)
Low dark current PMT
Luminescence Photon counting low dark current PMT
Plate formats
6 – 1536-well plates, NanoQuant Plate
Temperature control
Ambient +4 to 42 °C
Injection volume Variable; 1 µl increments
Syringe size Selectable: 500, 1000, 2500 µl
Pump speed 100 – 300 µl/sec
Dead volume 100 µl including pump back
Tutoring: First time users must order a tutoring session.
Booking
Description
Absorption, fluorescence, luminescence, and alpha- screen. The Infinite M1000 PRO offers broad wavelength selection, from UV to NIR, with flexible bandwidth adjustment for optimization of fluorescence intensity based applications.
Features/Specifications
Wavelength selection
Premium Quad4 Monochromators
Absorbance Fluorescence
Bandwidth Fixed, 5 nm Adjustable = 300 nm: 2.5 – 10 nm;
> 300 nm: 5 – 20 nm
Wavelength accuracy ± 0.3 nm = 300 nm: ± 0.5 nm;
> 300 nm: ± 1 nm
Wavelength reproducibility ± 0.3 nm = 300 nm: ± 0.5 nm;
> 300 nm: ± 1 nm
Wavelength range
Absorbance Fluorescence FP
230 – 1,000 nm Ex: 230 – 850 nm; Em: 280 – 850 nm Ex (LED):
470, 530, 590, 635 nm
Em (Mono):
280 – 850 nm
Detectors
Absorbance UV Silicon photodiode, 4-channel parallel reading
Fluorescence (incl. FP) Extended wavelength (UV and far red)
Low dark current PMT
Luminescence Photon counting low dark current PMT
Plate formats
6 – 1536-well plates, NanoQuant Plate
Temperature control
Ambient +4 to 42 °C
Injection volume Variable; 1 µl increments
Syringe size Selectable: 500, 1000, 2500 µl
Pump speed 100 – 300 µl/sec
Dead volume 100 µl including pump back
Tutoring: First time users must order a tutoring session.
Details
Name | Tecan infinite M1000 Pro |
---|---|
Manufacturers | Tecan |
×
Fingerprint